Early displacement of an atrial lead: A fatal presentation of a relatively frequent complication

Authors

  • Gala Caixal Parc Taulí Hospital Universitari – Cardiology Department – Arrhythmia Unit – Sabadell, Barcelona, Spain | Universitat Autònoma de Barcelona – Departament of Medicine – Barcelona, Spain. https://orcid.org/0000-0003-1830-3125
  • Victor Garcia-Hernando Parc Taulí Hospital Universitari – Cardiology Department – Arrhythmia Unit – Sabadell, Barcelona, Spain | Universitat Autònoma de Barcelona – Departament of Medicine – Barcelona, Spain. https://orcid.org/0000-0003-1190-6679
  • Jordi Punti Parc Taulí Hospital Universitari – Cardiology Department – Arrhythmia Unit – Sabadell, Barcelona, Spain | Universitat Autònoma de Barcelona – Departament of Medicine – Barcelona, Spain. https://orcid.org/0009-0002-1880-1706
  • Maria Soledad Ascoeta-Ortiz Parc Taulí Hospital Universitari – Cardiology Department – Arrhythmia Unit – Sabadell, Barcelona, Spain | Universitat Autònoma de Barcelona – Departament of Medicine – Barcelona, Spain. https://orcid.org/0000-0002-9455-5861
  • Gabriel Gusi Parc Taulí Hospital Universitari – Cardiology Department – Arrhythmia Unit – Sabadell, Barcelona, Spain | Universitat Autònoma de Barcelona – Departament of Medicine – Barcelona, Spain. https://orcid.org/0009-0006-9416-949X
  • Antonio Martinez-Rubio Parc Taulí Hospital Universitari – Cardiology Department – Arrhythmia Unit – Sabadell, Barcelona, Spain | Universitat Autònoma de Barcelona – Departament of Medicine – Barcelona, Spain. https://orcid.org/0000-0002-2070-4617

DOI:

https://doi.org/10.24207/jca.v38i1.3561

Keywords:

Dual chamber pacemaker, Atrial lead displacement, Ventricular arrythmia, Sudden cardiac death

Abstract

An 81-year-old woman was referred to Parc Taulí Hospital Universitari with complete atrioventricular block and a dual-chamber pacemaker was implanted. She subsequently presented an episode of ventricular tachycardia that degenerated into ventricular fibrillation requiring cardiopulmonary resuscitation. After review of the device, as well as the electrocardiogram and chest X-ray, displacement of the atrial lead into the right ventricle was confirmed. Contrary to expectations, the ventricular event was not secondary to endocardial irritation by the displaced lead, so the hypothesis of a possible depolarization-repolarization heterogeneity as a cause of the increased arrhythmogenicity was raised.

Downloads

Download data is not yet available.

References

Fuertes B, Toquero J, Arroyo-espliguero R, Lozano IF. Review Article Pacemaker Lead Displacement: Mechanisms And Management. Indian Pacing Electrophysiol J. 2003 [cited 1996 Jun 5];3(4):231–8. Available from: https://pubmed.ncbi.nlm.nih.gov/16943923/

Qin D, Filippaios A, Murphy J, Berg M, Lampert R, Schloss EJ, et al. Short- and Long-Term Risk of Lead Dislodgement Surveillance Registry. Circ Arrhythm Electrophysiol. 2022;15(8):e011029. https://doi.org/10.1161/circep.122.011029 DOI: https://doi.org/10.1161/CIRCEP.122.011029

Guan F, Li G, Liu Y, Gao X, Zhou R. Delayed management of atrial lead dislodgment after pacemaker implantation : a case report. J Med Case Rep. 2021;15(1):9. https://doi.org/10.1186/s13256-020-02626-z DOI: https://doi.org/10.1186/s13256-020-02626-z

Lin G, Zou T, Du M, Tong J, Shi H, Chen H, et al. Lead dislodgement complicated with paroxysmal atrial fibrillation as a rare cause of inappropriate implantable cardioverter defibrillator shock: a case report. J Geriatr Cardiol. 2021;18(10):851-854. https://doi.org/10.11909/j.issn.1671-5411.2021.10.002

Beer WM, Cherry DC, Goldenberg WD, Demers G. Dislodged pacer lead causing ventricular tachycardia. J Emerg Med. 2015;48(6):e145–6. https://doi.org/10.1016/j.jemermed.2014.12.075 DOI: https://doi.org/10.1016/j.jemermed.2014.12.075

Witt CM, Asirvatham SJ, Warnes CA, McLeod CJ. Ventricular tachycardia secondary to abandoned epicardial pacemaker lead. Hear Case Reports. 2015;1(3):126–9. https://doi.org/10.1016/j.hrcr.2015.01.009 DOI: https://doi.org/10.1016/j.hrcr.2015.01.009

Sharif AY, Davoodi G, Saeed AK, Sadeghian S. New technique: Repositioning of dislodged atrial pacing lead with a specially designed urological basket. Europace. 2007;9(2):105–7. https://doi.org/10.1093/europace/eul143 DOI: https://doi.org/10.1093/europace/eul143

Afzal MR, Horner S, Matre NB, Blake P, Dunham K, Pinkhas D, et al. Comprehensive strategy to reduce the incidence of lead dislodgement for cardiac implantable electronic devices. Pacing Clin Electrophysiol. 2019;42(1):58–62. https://doi.org/10.1111/pace.13544 DOI: https://doi.org/10.1111/pace.13544

Downloads

Published

2025-03-24

How to Cite

Caixal, G., Garcia-Hernando, V., Punti, J., Ascoeta-Ortiz, M. S., Gusi, G., & Martinez-Rubio, A. (2025). Early displacement of an atrial lead: A fatal presentation of a relatively frequent complication. JOURNAL OF CARDIAC ARRHYTHMIAS, 38. https://doi.org/10.24207/jca.v38i1.3561

Issue

Section

Cardiac Pacing